Features of ESS Chamber:
Different stress can be set to screen the temperature change rate:5℃/min、10℃/min、15℃/min、20℃/min. Meet the stress screening and lead-free process of electronic equipment products. Wide temperature range from -70℃~+180℃
Application of ESS Chamber:
electronic equipment products
Testing standards:
IEC 60068-2-1 IEC 60068-2-2: 2007 IEC 60068-2-14 IEC 60068-2-30 IEC 60068-2-38 IEC 60068-2-78 ISO 16750-4 MIL-STD-810G ASTM E1457-13 JESD22-A10 MIL-STD-2164、MIL-344A-4-16、MIL-2164A-19、NABMAT-9492、GJB-1032-90、GJB/Z34-5.1.6、IPC-9701
Technical Parameters
| Temp & Humidity | SZ-A-FX-225C/D/G、SZ-A-FX-408C/D/G、SZ-A-FX-608C/D/G、SZ-A-FX-1000C/D/G | |
| Temp only | SZ-D-FX-225C/D/G、SZ-D-FX-408C/D/G、SZ-D-FX-608C/D/G、SZ-D-FX-1000C/D/G | |
| Temp range | C: -20℃~+150/+180℃ D: -40℃~+150/+180℃ G: -70℃~+150/+180℃ | |
| Temp control accuracy | Temp resolution: 0.01℃ Temp fluctuation: ≤±0.5℃ Temp variation: ≤±2.0℃ Temp gradient: ≤±2.0℃ | |
| Heating/Cooling rate | F5≥5℃/min F10≥10℃/min F15≥15℃/min F20≥20℃/min (Linear/Nonlinear) | |
| Humidity range | 20%~98%RH (by +20℃~+85℃) standard 10%~98%RH (by +20℃~+85℃) optional | |
| Humidity control accuracy | Humidity resolution: 0.1%RH Humidity fluctuation: ≤±3%RH Humidity variation: ≤±3%RH (>75%RH); ≤±5%RH (≤75%RH) | |
| Heater | Nichrome electronic heater | |
| Refrigeration unit | Refrigeration method: Air cooling/ water cooling | |
| Workspace dimensions (W*H*D) | 500*750*600、600*850*800、800*950*800、1000*1000*1000 | |
| Exterior dimensions (W*H*D) | 1000*1750*1300、1100*1850*1500、1300*1950*1500、1500*2000*1700 | |